single-rb.php

JRM Vol.9 No.2 pp. 140-145
doi: 10.20965/jrm.1997.p0140
(1997)

Paper:

Uncertainty Analysis of Three-Dimensional Coordinate Measuring Machines

Jiro Matsuda

National Research Laboratory of Meteorology, 1-1-4 Umezono, Tsukuba 305, Japan

Received:
November 13, 1996
Accepted:
December 20, 1996
Published:
April 20, 1997
Keywords:
Uncertainty, Coordinate measuring machine step gauge, Analysis variance, Trigger probe
Abstract
On the basis of the condition that the distances between planes be measured in accordance with the ISO guide "Uncertainty of Measurement", the uncertainty of measurement of a CMM in the orthogonal coordinate system has been obtained by carrying out a systematic experiment. So the expanded uncertainty is given as 4.4μm.
Cite this article as:
J. Matsuda, “Uncertainty Analysis of Three-Dimensional Coordinate Measuring Machines,” J. Robot. Mechatron., Vol.9 No.2, pp. 140-145, 1997.
Data files:

*This site is desgined based on HTML5 and CSS3 for modern browsers, e.g. Chrome, Firefox, Safari, Edge, Opera.

Last updated on Dec. 06, 2024